Capstar-100 C02 Analyzer

The CAPSTAR-100 Carbon Dioxide Analyzer provides continuous measurement of expired CO2 over the range 0-10 percent. Its rapid response time, small sample flow requirements, and long-term stability make it ideal for respiratory gas measurement of rodents and larger animals.

The heart of the instrument is a miniature infrared analyzer with exceptional stability, low maintenance, and a linear output. The heated measurement cell prevents water condensation. The SENSORPOD, which contains the infrared measurement cell, attaches to the front panel by a 1.5m cable. This cell can be removed for cleaning, inspection, or replacement in minutes.

The digital display shows CO2 concentration in percent or mmHg, and allows breath-by-breath peak end-tidal reading or continuous rapid response. Calibration is performed with a single calibrated gas and room air.

The reliability and ease of operation of the CAPSTAR-100 make it the perfect instrument for routine respiratory CO2 monitoring. It is the natural companion to our SAR-830 series Small Animal Ventilators for monitoring respiratory status.

With the AVS-1 Advanced Ventilator System, it can control end-tidal CO2. It can also be used to monitor unassisted animals. Gas sampling rates of less than 30ml/min can be used, making it possible to monitor animals as small as rats. The fast response time of the CAPSTAR-100 allows end-tidal response plateaus with respiratory rates of 120 breaths/min or greater.

A complete range of accessories is available to ensure easy setup and convenient operation. An accessory pack containing sample tubing, moisture traps, and Nafion® moisture- absorbing sample tubes is included with the instrument.



Analyzer type
Sample cell materials
Sample inlet connecto
Recommended sample tubing size
Measurement range
Analog output scaling
Linearity
Repeatability
Response time (T10 - T90)
Response time (T10 - T90)
Response time (T10 - T90)
Zero stability
Operating temperature range
Optical bench temperature
Warm-up time
Sample rate flowmeter
Maximum sample cell pressure anesthetic agents
Interference effects: 50% N2O
Nitrous oxide compensation
Controls
Electrical requirements
Dimensions (main unit)
Dimensions (SENSORPOD)
Weight
single beam infrared
.sapphire and stainless steel
0.062” OD brass tube
0.040" - 0.062" ID Tygon™
0-10% CO2
100mV/% CO2 (0-1V)
.0.1% CO2
0.1% CO2
100mS at 100 ml/min sampling
130mS at 50 ml/min sampling
250mS at 30 ml/min sampling
0.1% (8 hours), 0.2% (24 hours)
5 - 40oC
48oC, temperature controlled
5 min to 0.2%, 10 min to 0.1%
10 - 50 ml/min
.+5 psig
negligible
0.1% at 0% CO2, 0.6% at 5% CO2 (uncompensated)
.0.1% at 5% CO2
Zero, Span (10-turn dials)
120VAC/220VAC switchable, 35VA
9W x 5H x 9D in., 23W x 13H x 23D cm
3W x 2.5H x 3D in., 7.6W x 6.4H x 7.6D cm
10 lbs (4.5kg)

Ordering Information
11-10000 CAPSTAR-100 CO2 Analyzer, complete with accessory kit , 120/240V 50/60Hz
11-01100 Replacement CO2 sample cell
11-01101 Accessory kit; 2 moisture traps, 2 13mm hydrophobic filters,
1 sample line, 2 Nafion®. sample lines (6” and 2’)
11-01102 Sample line, plain, 5', with male Luer-loc connectors
11-01103 Sample line, 2' Nafion® tubing (absorbs and evaporates condensed moisture), with Luer connectors
11-01111 Sample line, 6” Nafion® tubing, with Luer connectors
11-01104 Sample tubing, .040" ID Tygon™, 10' coil
11-01105 Luer-loc adapter, .040" - .062" ID flex tubing to male Luer, pack of 5
11-01106 Luer-loc adapter, .125" ID flex tubing to male Luer, pack of 5
11-01107 Moisture trap, male-female Luer-loc connectors
11-01108 In-line miniature (13mm) hydrophobic filter, Luer male-female fittings
11-01110 Calibration gas, 4 liter aerosol can, 5% CO2, 20% O2, balance N2


IITC Life Science Inc.
23924 Victory Blvd Woodland Hills, CA 91367
818 710 8843 fax 818 992 5185
 
  Copyright (c) 2007. IITC Life Science
All rights reserved.